Single Pump Hemodialysis Machine
Technical Parameters
Extracorporeal circulation system
Blood pump flow: 10~620ml/min
Blood pump segment Dia.: 6.4 and 8mm adjustable
Heparin pump flow scope: 0~10ml/h
Syringe: 10ml/20ml/30ml;20ml/30ml/50(optional)
Arterial and pressure monitor: -300~+400mmHg(±10mmHg)
Venous pressure monitor: -50~+300mmHg(±10mmHg)
Air bubble detector: Infrared and ultrasound detection
Dialysis system
Dialysate flow: 300~800ml/min linearly adjustable(+10%,-5%)
Dialysate temperature scope: 35.0~39.0ºC
Conductivity: 12.3~15.8Ms/cm
Na+ adjustment scope: 125~151 mmol/L
HCO3- adjustment scope: 24.0~40.0 mmol/L
Bicarb Cartridge: Optional
Blood leakage detection
Method: Photoelectricity detection
Ultrafiltration controlling system
Ultrafiltration ratio: 0~4000ml/h
Pump volume accuracy: ±1%
Disinfection and rinse
Rinse and heat disinfection(above 90ºC)
Chemical disinfection: Peracetic acid
Heat chemical disinfection: Citric acid
Other parameter
Dimension: Length*Width*Height:370mm*340mm81450mm
Weight: 90Kg
Power supply voltage: 100~240V Ac±10% 50~60HZ
Power: 1.7KW
Back-up battery: At least 30 minutes working after electricity cut
Working environment: Temperature: 10~30ºC
Relative humidity no higher than 70%
Water supply temperature: Temperature: 5~30ºC
Dialysate formula: Great compatibility be applicable for various concentration ratio formula
Pressure: 0.12~0.6Mpa